Applies to: Z4

Starting the machine

  1. Make sure that the machine is properly installed.

  2. Switch on the machine at the main power switch.

  3. Press the start button.

  4. Launch the dentalcam computer.

  5. Start dentalcnc.

    • The working chamber lighting is on. The machine references.

  6. The machine will not reference if the working chamber door is open.

  7. If the working chamber door is open, close it.

Starting the machine with a tool in the collet chuck

Under some circumstances, such as a power failure, there can be a tool in the collet chuck of the spindle when you start the machine.

NOTICE

Machine damage if you do not remove the tool

If the tool remains in the spindle after confirmation of the message, it will collide with and damage machine parts.

  • Remove the tool as described below.
  1. Start the machine.
    • dentalcnc displays that there is a tool in the collet chuck. The working chamber door opens.
  2. CAUTION!Wear gloves.

  3. Hold the tool in the collet chuck in place.

  4. Confirm the current message.
    • The collet chuck opens.
  5. Remove the tool from the collet chuck.
  6. Confirm the current message.
    • The working chamber door closes. The machine is ready for operation.