Commissioning
Watch the video
YouTube video – When viewing this video, personal data is sent to YouTube, LLC, USA. Privacy statement
The installation steps up to operation are:
Step 1: Removing the transport lock
Short-circuit hazard when the machine is too cold
If the machine is transported from a cold environment into a warmer environment, a short circuit may occur caused by condensate.
- Before switching on the machine after transportation, ensure the following:
- The ambient air has the allowed temperature.
- The machine has the same temperature as the ambient air. This will take at least 48 hours.
- The machine is completely dry.
Before operating the machine for the first time, you must remove the transport lock in the working chamber. The transport lock prevents the spindle from getting damaged during transport.
- Ensure the following:
- The machine is connected to the electrical source.
- The dentalcam computer is not connected to the machine.
- Switch on the machine at the main power switch.
- Open the working chamber door.
- Switch off the machine at the main power switch.
- Remove the transport lock as shown in the supplement.
-
Keep all parts of the transport lock for possible servicing returns.
Step 2: Connect to the dentalcam computer
ATTENTION! The transport lock must not be installed when you connect the CAM computer
The computer that is connected to your machine and which runs dentalcam and dentalcnc. to the machine.
-
Switch on the machine at the main power switch.
-
Close the working chamber door.
-
Connect the USB port on the machine's connection panel to the USB port on the dentalcnc computer.
-
Install the newest version of dentalcnc that is released for the machine.
-
dentalcnc 9
Set up the software connection to the machine.
(missing or bad snippet)dentalcnc 8
To determine the USB port number and connect to the machine, select the depicted icon in the Application settings. -
The machine references.
To avoid connecting errors, we highly recommend that you use the USB or network cable which was provided with the device. Original spare parts are available from customer service.
What's next?
If you want to create a job in dentalcam:
If you want to mount blanks: